The instructions for adjusting the injector for leakage are in section 9 and fig. 9 of the attached manual. When reattaching the handle after the adjustment, make certain that the set screws are engaging the flat sides of the shaft. When the injector is in the "inject" position, the handle should be pointed down and to the right when viewed from the front.