ENO-20/30

A large negative peak before nitrite is interfering with the nitrite peak with biological samples, what causes this?
The negative peak is caused by the change in matrix from the carrier to the sample. There is almost always a negative peak and this is normal. However, an a...
Fri, 4 Oct, 2019 at 2:25 PM
ENO-20 does not start. Sequencer Error
The ENO-20 has a rechargable battery inside to backup memory. If you do not plug in the system for more than a year, the battery will drain and the ENO-20 w...
Tue, 16 Jul, 2019 at 3:39 PM
ENO-20/30 NOx Analyzer shows Error 3
The ENO-20 gives error 3 when its pressure is unstable. There are two sensors on the pump, pressure sensor and rotating sensor of the pulley. If one of the...
Fri, 4 Oct, 2019 at 2:26 PM
ENO-20 restoration
If the ENO-20 has not been used for awhile without proper storage practice as shown in the manual, please treat as follows:   1. Turn on the system and co...
Tue, 16 Jul, 2019 at 3:45 PM
ENO-20/30 Shortened Retention times for tissue or blood Samples
If peak retention times decrease quickly even when using a newly packed  pre-column after 10-20 samples, the protein concentration may be too high. If you f...
Fri, 4 Oct, 2019 at 2:27 PM
ENO-20/30, Smaller Peak (small peak, less signal, weak signal) with Standard for Nitrite and Nitrate
1. Inject 10 µl of 10 µM standard from the manual injector, mix nitrite and nitrate. When using the manual injector, please set the autosampler injector at ...
Fri, 4 Oct, 2019 at 2:28 PM
ENO-20/30: How long can the Carrier/Reactor solution be stored, refrigerated before it is no longer good to use?
Carrier Solution: When carrying out high-sensitivity analysis, such as below 100 nM, the Carrier solution should be used within one week of preparation. F...
Thu, 18 Jul, 2019 at 10:04 AM
ENO-30 standard pressure
Carrier 1.5 MPa Reactor 1.2 MPa This is just for reference value. Higher or lower pressure does not directly mean a problem.
Thu, 18 Jul, 2019 at 10:05 AM
Frequent bubbles in pumps on ENO-20/30
One source of air bubbles is loose connections to the degasser inlet/outlets. The end of the tubing and flat side of the ferrule should be perfectly aligned...
Thu, 18 Jul, 2019 at 2:05 PM
how to inject samples from manual injector when AS-700 autosampler is set up on ENO-20/30
Please set the manual injector in the upstream of autosampler. When you want to inject sample or standard from the manual injector, please turn the autosampl...
Thu, 18 Jul, 2019 at 2:09 PM