Tomy Autoclave Solutions- SX Series
If an error code is displayed, please see the following table outlining the codes and and recommended action. Search terms: E01 E04 E05 E06 E07 E08 E09...
Thu, 7 Oct, 2021 at 10:08 AM
Nothing is displayed on the control panel, even with the power supply switch turned ON. Please check if the power cable is connected, or if the power suppl...
Thu, 7 Oct, 2021 at 10:09 AM
If the temperature rises very slowly, please check the volume of sterilizing water you are using. If you overfill the sterilizing water, it takes more time...
Thu, 7 Oct, 2021 at 10:09 AM
If there is a problem with sterilization, the sterilization time is insufficient, especially for larger volumes of liquid. Please increase the sterilizing ...
Thu, 7 Oct, 2021 at 10:10 AM
Please check the chamber lid gasket or chamber opening for dust or obstructions. Inspect the chamber lid gasket for damage or wear as it may be time to rep...
Thu, 7 Oct, 2021 at 10:11 AM
In order for the Tomy Autoclave to open, please ensure it is plugged in and turned on. Before you can open the lid, the LED lights need to be displayed. Th...
Thu, 7 Oct, 2021 at 10:12 AM
If steam is leaking from the safety valve, it means the pressure in the chamber is too high and further action including repair is required. Please contact ...
Thu, 7 Oct, 2021 at 10:13 AM
The set temperature range is as follows: Sterilizing: 105-135°C Heating: 45-104°C Warming: 45-95°C Additional support: Contact us Back to the To...
Thu, 7 Oct, 2021 at 10:13 AM
The preset timer allows start time delay up to 99 hours, allowing you to sterilize at the most convenient time. Sterilization function (105 – 135°C) progra...
Thu, 7 Oct, 2021 at 10:14 AM
The SX-500 holds 50 Liters, the SX-700 holds 69 Liters total volume. The SX-700 can accommodate up to 12 pieces of 1000 mL flasks or 27 pieces of 300 mL b...
Thu, 7 Oct, 2021 at 10:15 AM